Feed Roller Nip Drive Belt - Konica Minolta DRYPRO SIGMA Service Manual

Medical printer
Hide thumbs Also See for DRYPRO SIGMA:
Table of Contents

Advertisement

SERVICE MANUAL

Feed Roller Nip Drive Belt

Prerequisites:
Removal
Important
The belt and pulleys must be removed as a set for the belt to fit onto the pulleys.
Nip Drive Belt
and Pulleys
HC
[1] Remove the:
(a) E-rings from the 2 nip drive belt pulleys
(b) 2 pulleys from their shafts
(c) Nip drive belt
Installation
Reverse the steps in the removal procedure.
2-28
• Close the film cartridge. (Touch the Pause button on the display screen.)
• De-energize the laser imager.
• Disconnect the power cord.
Remove the right cover
9G9002 – 2011-08-15