Condition Code: 21-146 – Unsupported Media Size
Web Portal Message
Film Supply:
Unsupported Film Size
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Logged for RMS
Condition Code: 21-175 – Rollback Failed to Engage Cartridge
Web Portal Message
Rollback Failed to
Engage Cartridge
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Logged for RMS
9G9002 – 2011-08-15
Log Message
Film Supply: Media
Size in film cartridge is
not supported
The imager obtains the media size information from the RF Tag when the film
cartridge is loaded or when the imager is powered on. This error occurs if the
imager does not support the size.
The error is cleared when a film cartridge is removed and then re-inserted.
Cannot print from the loaded film cartridge. The spooler is still able to accept
and queue jobs. [SI-2]
The laser imager has not been configured to use this film size.
1. Try a different film size or reconfigure the laser imager.
2. If the error persists, call for service.
True
WARNING
Unique Symb. ID: ERR_UP_WRONG_MEDIA_SIZE
[CartridgeInfo=bbeeee-cjj-d-iii][FilmSize=size]
Log Message
Rollback failed to
engage cartridge
This error occurs if the rollback mechanism reactivates at the home position
(S2x) because the rollback mechanism is rotating in place. The rollback motor
is moving (because the home sensor changes state), but it does not travel
toward the open sensor. The most likely cause is a mis-inserted cartridge.
The error is cleared when the film supply is closed (as indicated by the supply
drawer interlock switch i1x).
Cannot print from the loaded film cartridge. The spooler is still able to accept
and queue jobs. [SI-2].
1. Insert a new film cartridge.
2. If the error persists, call for service.
True
WARNING
ERR_UP_ROLLBACK_ENGAGE
[CartridgeInfo=bbeeee-cjj-d-iii][FilmSize=size]
DICOM Printer
StatusInfo
BAD SUPPLY MGZ
DICOM Printer
StatusInfo
BAD SUPPLY MGZ
Diagnostics
Severity
1-Alert
Severity
1-Alert
4-53