Konica Minolta DRYPRO SIGMA Service Manual page 118

Medical printer
Hide thumbs Also See for DRYPRO SIGMA:
Table of Contents

Advertisement

SERVICE MANUAL
Condition Code: 20-209 – Imager Opened During Self Test
Web Portal Message
Laser Imager Opened
During Self Test
Description
System Impact
User Action
FE Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Condition Code: 20-449 – Processor Filter Maintenance Required
Web Portal Message
None
Description
System Impact
User Action
FE Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
4-40
Log Message
Top, Side, or Front
Cover Opened During
Self Test
This error occurs during MCS diagnostics when the laser imager determines
that a cover was opened during the execution of self test. The error clears only
when power is cycled.
The imager is unable to print films. The spooler is still able to accept jobs but
new films will not be initiated. Films in progress will not be completed. [SI-3]
1. Close the open top cover or the open cover.
2. Shut down the laser imager, then restart the laser imager.
3. If the error persists, call for service.
1. Check that the following are closed and secured:
• Top cover
• Film supply (front) and back covers
2. De-energize and then energize the laser imager.
True
FAILURE
ERR_MCS_DOOR_OPENED_DURING_SELF_TEST
Log Message
Processor Filter
Maintenance Required
This notification is written to the log when the processor filter maintenance
(PFM) sheet count exceeds the PFM sheet count limit.
Note
This notification is shown to the user but is different from other "errors" because
it is a special situation.
None. The system will function normally. [SI-0]
None
1. Change the filter.
2. Use the service tool to reset the Counts to Filter Change.
True
NORMAL
ERR_PFM_REQUIRED
DICOM Printer
StatusInfo
PRINTER DOWN
DICOM Printer
StatusInfo
NORMAL
Severity
0-Fatal
Severity
3-Notify
9G9002 – 2011-08-15

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents