Caution
Isopropyl alcohol is highly flammable. Follow the manufacturer's instructions for safe use and handling.
[19] Use alcohol wipes and a scraper to inspect and clean debris and buildup from the:
• Heat shield, including the bottom side of the shield
• Cooling plate
• Densitometer guide
Densitometer
Guide
[20] Vacuum the:
• Top cover
• Film supply
• Entire processor assembly
[21] Install a new felt pad assembly with 2 screws (see Adjustments and Replacements,
[22] Carefully remove the film you inserted over the imaging roller and discard it.
[23] Close the heat shield.
[24] Install the 2 rear screws on the processor drum cover.
[25] Install the:
• New deodorant filter
• Right cover
• Left cover
• Top cover
[26] Close the film supply.
[27] Place all used alcohol wipes and cleaning pads in a bag, close the bag with a tie-wrap, and then dispose of it
according to local ordinances.
[28] Energize the imager.
[29] Perform
"Resetting the PM Counts and Changing the Service
9G9002 – 2011-08-15
Cooling Plate
Heat Shield
"Felt Pad
History".
Preventive Maintenance
Assembly").
3-7