SERVICE MANUAL
Condition Code: 01-000 – General Configuration and Control Subsystem Error
Web Portal Message
None
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Condition Code: 01-001 – Maximum Number of Allowed Restarts for a Subsystem or
Component Has Been Exceeded
Web Portal Message
None
Description
System Impact
User Action
DICOM Printer
Status
Unique Symb. ID
4-10
Log Message
General Error in
Configuration and Control
Subsystem
A general error occurred in the Configuration and Control Subsystem.
Unknown. This is an unexpected or unknown error; it is not possible to
document specific details.
None
False
NORMAL
ERR_CONFIG_CONTROL
Log Message
Maximum Number of
Restarts Exceeded for xxxx
(where xxxx is the
subsystem name)
When a subsystem or component generates a non-recoverable, Configuration-
and-Control will attempt to shut down and restart it. This error means that
Configuration-and-Control has initiated more than the configured number of
allowed restarts.
The system will continue to run normally, unless the subsystem or component
in question cannot run any longer.
None
NORMAL
ERR_EXCEED_MAX_RESTARTS
DICOM Printer
StatusInfo
NORMAL
3-Notify
DICOM Printer
StatusInfo
NORMAL
2-Warning
Severity
Severity
9G9002 – 2011-08-15