Konica Minolta DRYPRO SIGMA Service Manual page 121

Medical printer
Hide thumbs Also See for DRYPRO SIGMA:
Table of Contents

Advertisement

Condition Code: 20-705 – Restart in Process
Web Portal Message
None
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Condition Code: 20-706 – Shutdown Completed
Web Portal Message
None
Description
System Impact
User Action
FE Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
9G9002 – 2011-08-15
Log Message
None
This auxiliary code indicates that a user-initiated system restart is in process.
The user should not power the imager off during the startup sequence but wait
for the sequence to complete. The code clears after the sequence completes.
Note
One of the reasons that the operating system would be restarted is for a software
update procedure.
The imager is unable to print films.
None
True
N/A
ATTEN_PRINTERINFO_RESTART
Note
This is a printer status indication, not an error condition.
Log Message
None
This auxiliary code indicates that a user-initiated system shutdown is complete.
(The shutdown would have been initiated remotely from the Web Portal
because shutting down the imager at the machine accomplishes the power-off
without initiating a shutdown.) The imager must now be powered off using the
power switch. When this code is displayed, the imager can be restarted only by
using the power switch to power off and then power on.
The imager is unable to print films.
None
De-energize the laser imager.
True
N/A
ATTEN_PRINTERINFO_SHUTDOWN
Note
This is a printer status indication, not an error condition.
DICOM Printer
StatusInfo
N/A
DICOM Printer
StatusInfo
N/A
Diagnostics
Severity
4-Auxiliary
Severity
4-Auxiliary
4-43

Hide quick links:

Advertisement

Table of Contents
loading