SERVICE MANUAL
Condition Code: 04-110 – Invalid Association Request
Web Portal Message
None
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Condition Code: 04-200 – Disk Full
Web Portal Message
MIM Core: Disk Full
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
4-16
Log Message
The association
request was rejected
An association request was rejected for one or more of the following reasons:
• The Service Class User (SCU) did not specify an implementation UID.
• The SCU did not specify an application context name.
• The protocol version is a two-byte field; bit zero (0) must be one (1).
• The SCU did not specify any presentation context items to be negotiated.
• None of the proposed presentation context items were successfully
negotiated. This may happen, for example, when a DICOM Store SCU tries
to communicate with a DICOM Print SCP.
The SCU will not be able to open an association, and therefore will not be able
to deliver a print job.
None
False
NORMAL
ERR_INVALID_ASSOCIATION
Note
Rationale for not reporting to user: This is a DICOM SCU configuration problem,
not an operational error. The user cannot correct this from the imager.
Additional DICOM information: There is a specific result/reason code available
for this circumstance: 1-Rejected permanent/7-Called AE title not recognized.
Log Message
Disk Full
The partition of the disk where studies are stored is full.
If a print request is in progress, it cannot be completed. New studies will not be
accepted.
1. Load the requested film type for jobs waiting to be printed.
2. If the error persists, call for service.