Condition Code: 06-500 – Internal Page Delivery Component Error
Web Portal Message
None
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
4-32
Log Message
• BIND error. Code =
YY; Page ID = XXX
• STAGE error. Code
= YY; Page ID =
XXX
• PRINT error. Code
= YY; Page ID =
XXX
• EXPOSE error.
Code = YY; Page
ID = XXX
• Unknown error.
Code = YY; Page
ID = XXX
One of the following internal errors occurred:
• Binding error
• Staging error
• Printing error
• Exposure error
• Unknown internal error
The job may not be delivered due to an internal error communicating with the
MCS Interface Server (MIS).
None
False
NORMAL
ERR_PAGE_DELIV_COMP
Note
The log text will reflect the specific cause from the listed set of causes.
Rationale for not reporting to user: This is an error on an individual job and does
not affect imager operation. THe imagerwill automatically attempt to printthe job
again.
DICOM Printer
StatusInfo
NORMAL
Severity
1-Alert
9G9002 – 2011-08-15