Section 2: Adjustments and Replacements
Processor—Drum Temperature Adjustment
Adjustment Specification
To set the correct temperature for the drum
Purpose:
Setting Temperature: This procedure is required after replacing the processor assembly, processor drum, or the
DRE board, or if you expect that drum temperatures are causing image quality problems.
The temperature at the processor drum must be between 123.5–124.5 °C.
Specification:
Special Tools:
None
Prerequisites
Note
• A temperature meter and probe with a bar type element must be used for this procedure. Probes with circular
type elements will not provide correct readings. The temperature meter and probe must be calibrated together,
at least once per year. If the probe breaks, a new probe and the existing meter must be calibrated again.
• The temperature meter must be at room temperature during this procedure. If the meter has been in a hot or cold
vehicle, allow it to adjust to room temperature before using.
9G9002 – 2011-08-15
• Temperature meter with probe
• Block for probe, kit
• Laptop computer with MICROSOFT INTERNET EXPLORER
• Service tool
• Crossover cable

Adjustments and Replacements

2-1