Condition Code: 06-420 – Invalid Data or Data Access Error
Web Portal Message
MIM Core: Internal
Software Error
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
9G9002 – 2011-08-15
Log Message
• Page width and/or height are
less than zero
• Invalid page band
• Invalid coordinates for
indicator icon: (X1,Y2),
(X2,Y2). (where X1, Y1 and
X2, Y2 are the calculated
icon coordinates)
• Image size mismatch error
• Error in cropping coordinates
• The file XXXX could not be
found (where XXXX is the file
name)
• Error reading image XXXX
(where XXXX is the name of
the image file)
• General error accessing
database components
• General error accessing
page formatter
The page builder component has encountered the following data error:
• The specified page width and/or height is less than zero.
• The number of page bands is neither 1 nor 3, and each of the bands does
not have the same type.
• Icon (x1,y1) and/or (x2,y2) is beyond the page extents.
• Image size mismatch error, where the size read does not match the size
expected.
• Bad cropping coordinates.
• Icon or image file not found.
• Trying to read image data; most likely, the image input file could not be
found.
• General error accessing the database component.
The current page will not be processed because it c ould not be rendered due to
an internal error.
1. Resend the print job from the image source.
2. If the error persists, call for service.
True
NORMAL
ERR_INVALID_DATA
Note
The log entry reflects the specific cause from the listed set of causes.
Current page will not be delivered. (Undeliverable)