Konica Minolta DRYPRO SIGMA Service Manual page 38

Medical printer
Hide thumbs Also See for DRYPRO SIGMA:
Table of Contents

Advertisement

SERVICE MANUAL
[5] Remove the drum drive belt from the exit pulley.
Exit Pulley
[6] To remove the belt completely, you must free it from the processor drum pulley:
(a) Open the wire saddle and disconnect the drum drive connector.
(b) Remove the front screw and loosen the rear screw on the drum drive bracket.
(c) Pivot the drum drive bracket toward the back of the imager and remove the belt from the pulley.
Installation
[1] Slide the belt over the processor drum pulley.
[2] Pivot the drum drive bracket back into place. Make sure that the C-shaped feature on the slip ring assembly
straddles the bracket.
[3] Install and tighten the 2 bracket screws.
[4] Connect the drum drive connector.
[5] Hold the tensioner in position and place the belt over the exit pulley and the tensioner pulley.
• The bottom of the belt sits on the top of the tensioner pulley.
• The top of the belt runs just underneath the top of the tensioner bracket.
[6] Set the tensioner onto the end of the screw and attach the spring.
[7] Slide the tensioner completely into place, and then install the 1 nut and 2 screws.
[8] Connect the connector to the drum drive motor.
[9] Install the covers.
Note
The teeth on the belt must mesh properly with the teeth on the pulley. The longest span of the belt should deflect a
maximum of 1.27 cm (0.50 in.) when finger pressure is applied.
2-18
Drum Drive Belt
Processor Drum
Pulley
Slip Ring Assembly
Wire Saddle
Drum Drive
Connector
Drum Drive
Bracket Screws
9G9002 – 2011-08-15