SERVICE MANUAL
Upgrading the Firmware in the Microcontrollers
Important
You must perform this procedure when:
• The imager is ghosted and there are changes for the microcontrollers
• Condition code 20-156 is generated
[1] Log on to the service tool (see
[2] Select Diagnostics > Scripts.
[3] Select MicroUpdate_WithAutoRestart_ST.vbs from the Script Name drop-down list.
[4] Click Run.
Note
This procedure updates the microcontrollers as necessary and then restarts the imager. Be patient, as this process
could take approximately 10 minutes.