SERVICE MANUAL
Condition Code: 27-651 – Exposure timeout notification
Web Portal Message
None
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Condition Code: 28-501 – Processor RTD Short Circuit
Web Portal Message
Processor: Internal
Hardware Failure
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Logged for RMS
4-68
Log Message
Image Exposure timed
out, job will be reprinted
This notification occurs when the image exposure does not complete in the
required time. The faulty image prints. The job fails and the imager attempts to
print the image again.
If this problem is detected on three successive films, error 27-650 is declared.
None. The system will function normally. [SI-0]
None
False
NORMAL
ERR_EXPOSE_TMO_RETRY
Log Message
Short in RTD circuit
This error can occur for many electromechanical reasons: Resistance
Temperature Detector (RTD) and connector leads touching, drum-slip-ring
failure, condensation (cold machine in warm environment), failed component in
analog-to-digital conversion circuitry on the processor board, etc.
The error clears only when imager power is cycled.
The imager can only print jobs that were in process when the error occurred. The
spooler can still accept jobs, but the jobs will not print until the error is resolved.
All jobs that were in process when the error occurred will reprint completely
when the error is resolved. [SI-1]
1. Shut down the laser imager, and then restart the imager.
2. If the error persists, call for service.
True
FAILURE
ERR_PROCESSOR_RTD
[TriggerValues=a, b][TriggerNotes=RTD zone number, temperature in
degrees C]
DICOM Printer
StatusInfo
NORMAL
DICOM Printer
StatusInfo
PROC DOWN
Severity
3-Notify
Severity
0-Fatal
9G9002 – 2011-08-15