Condition Code: 21-632 – Dmax Not Met
Web Portal Message
Film Supply: Film
Calibration Failure -
Dmax Outside Target
Description
System Impact
User Action
FE Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Logged for RMS
9G9002 – 2011-08-15
Log Message
Film Supply:
Calibration failed -
Dmax Not Met
The error occurs when the calibration sheet does not have a step density at or
above the required Dmax value. The most likely causes are expired film or
incorrect processor temperature. The error is cleared when a new calibration is
initiated or when the film cartridge is removed. Any cartridge using these
calibration results will automatically go into AIQC-Manual Mode and will
continue printing but with reduced range in achievable densities. A new
calibration will be attempted when this cartridge is re-inserted.
None. The system will function normally. [SI-0]
The maximum density of the film is too low. Calibration results for this film
are outside the normal range but will still be used.
If you do not wish to use these calibration results, perform a new calibration
on this film or try another film cartridge.
Use the service tool to check the processor temperature and set point for
correct values. If necessary, adjust the temperature (see Adjustments and
Replacements,
"Processor—Drum Temperature
True
WARNING
ERR_UP_DMAX_NOT_MET
[CartridgeInfo=bbeeee-cjj-d-iii][FilmSize=size][TriggerValues=a,b,c,...,u]
[TriggerNotes=Optical Density values, from max to min]
DICOM Printer
StatusInfo
CALIBRATION ERR
Adjustment").