[4] Insert the probe into the slot on the left rear of the processor assembly, feed it upward, and let the rotation of
the processor drum pull the probe until the block reaches the slot.
Note
If the laser imager is in Standby mode, the processor drum pulls the probe very slowly and it may be difficult to detect
movement.
Slot
Probe
[5] Allow the temperature reading on the meter to stabilize. This requires approximately 90 seconds.
[6] Check that the temperature at the processor drum is between 123.5–124.5 °C.
[7] Pull back slightly, 1.0–2.0 cm (0.4–0.8 in.), on the block to confirm that the highest temperature is being read.
[8] Is the temperature within specification?
Yes
Remove the probe, and then
install the back cover.
9G9002 – 2011-08-15
Processor Assembly
No
Note the highest temperature
reading, and then perform the
temperature adjustment
procedure.