Optics And Exposure Transport - Konica Minolta DRYPRO SIGMA Service Manual

Medical printer
Hide thumbs Also See for DRYPRO SIGMA:
Table of Contents

Advertisement

SERVICE MANUAL

Optics and Exposure Transport

Remove the imaging
Prerequisites:
T20 Torx driver
Tools:
Removal
[1] Remove 2 screws on each side of the imaging assembly where the optics and exposure transport are joined.
[2] Carefully separate the optics from the transport.
Screws
Optics
Installation
Reverse the steps in the removal procedure.
Note
Use the alignment pins near the screw holes to help line up the two components.
2-40
assembly.
Exposure Transport
Screws
9G9002 – 2011-08-15