[2] Remove the 2 motor bracket screws.
[3] Hold the motor and bracket to keep the remaining connectors free of tension, and then remove the 2 screws
that attach the motor.
[4] If you need to keep the gear to use with the new motor, remove the:
(a) E-ring
(b) Gear from the shaft
Installation
Reverse the steps in the removal procedure.
9G9002 – 2011-08-15
E-ring
Gear
Bracket Screw
Bracket Screw
Adjustments and Replacements
Motor Screws
Motor
H221_6012HC