Section 3: Preventive Maintenance

Necessary Materials

The materials necessary for performing Preventive Maintenance (PM) are:
• Tools
• PM materials
Required Tools
• Laptop computer with MICROSOFT WINDOWS 98 operating system or higher and MICROSOFT INTERNET
EXPLORER
• Service tool
• Crossover cable
• Temperature meter with probe
• Block for probe, kit
• Vacuum cleaner
• Flashlight
• Scraper – recommended blade size is 1.25 × 4.75 in.
• T20 Torx driver
• T10 Torx driver
PM Materials
Part No.
A4A99G7543
A4A99G5904

PM Intervals

Important
A Preventive Maintenance (PM) call must be made every 20,000 film cycles. To consistently produce high quality
images, the imager must receive periodic routine maintenance. Perform PM procedures:
• Every Emergency Maintenance Call (EM)
• Every 20,000 cycles the display screen displays a message prompting the user to schedule preventive
maintenance. The procedures listed in the 20,000 cycle checklist should be performed as soon as possible after
the 20,000 cycle PM prompt.
9G9002 – 2011-08-15
Description
3M Auto-Pak tack cloth
Alcohol wipes
Cleaning pads, 10 × 10 cm (4 × 4 in.)
Bag with tie-wraps
Deodorant filter
Felt pad assembly
Safety glasses with side shields
Notes
Texwipe TX1065, 50 each
100 each
Use one bag for each PM.
Take 3 to the PM site
Replace every PM (20,000 cycles)