SERVICE MANUAL
SUPPLEMENTARY
INFORMATION
Repair information for distortion of index images

Contents

With 35mm film holder, the subject's slanting line in the 6th (or 4th) frame of index image may be distorted
like wave.
Service
Replace the Holder Hold SP with a Holder Hold SP (2892-0230) of service parts.
The Holder Hold SP replacement procedure
1. Remove two screws as shown, and remove the holder.
2. Rotate the motor axis in the arrow direction as shown until it stops.
3. Remove five screws as shown, and remove the base plate.
4. Remove one screw as shown, and replace the spring with Holder Hold SP (2892-0230) of service parts.
Base plate
Base plate
Base plate
Base plate
Base plate
Related information
:QS FA 2892-P002
This information should be filed in your service manual.
(2892)
CODE NO.
MODEL
H o l d e r
H o l d e r
H o l d e r
H o l d e r
H o l d e r
KONICA MINOLTA PHOTO IMAGING, INC.
1
No. QS FA 2892-S01E Jul..2005
2892-100
DiMAGE Scan Elite 5400 II
Holder Hold SP
Holder Hold SP
Holder Hold SP
Holder Hold SP
Holder Hold SP