Section 15: TSP command reference
trigger.generator[N].assert()
This function generates a trigger event.
Type
TSP-Link accessible
Function
Yes
Usage
trigger.generator[N].assert()
N
Details
Use this function to directly trigger events from the command interface or a script. For example, you
can trigger a sweep while the instrument is under script control.
Example
trigger.generator[2].assert()
Also see
trigger.generator[N].EVENT_ID
trigger.generator[N].EVENT_ID
This constant identifies the trigger event generated by the trigger event generator.
Type
TSP-Link accessible
Constant
Yes
Usage
eventID = trigger.generator[N].EVENT_ID
eventID
N
Details
This constant is an identification number that identifies events generated by this generator.
To have another trigger object respond to trigger events generated by this generator, set the other
object's stimulus attribute to the value of this constant.
Example
digio.trigger[5].stimulus = trigger.generator[2].EVENT_ID
Uses a trigger event on generator 2 to be the stimulus for digital I/O trigger line 5.
Also see
trigger.generator[N].assert()
15-368
Model 2606B System SourceMeter® Instrument Reference Manual
Affected by
Where saved
The generator number (1 or 2)
(on page 15-368)
Affected by
Where saved
The trigger event number
The generator number (1 or 2)
(on page 15-368)
Default value
Generates a trigger event on generator 2.
Default value
2606B-901-01 Rev. C November 2021