Model 2606B System SourceMeter® Instrument Reference Manual
smuX.measure.interval
This attribute sets the interval between multiple measurements.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
interval = smuX.measure.interval
smuX.measure.interval = interval
interval
X
Details
This attribute sets the time interval between measurements when smuX.measure.count is set to a
value greater than 1. The SMU attempts to start each measurement when scheduled. If the SMU
cannot keep up with the interval setting, measurements are made as quickly as possible.
If filtered measurements are being made, the time interval is from the start of the first measurement
for the filtered reading to the first measurement for a subsequent filtered reading. Extra
measurements made to satisfy a filtered reading are not paced by this interval.
Example
smua.measure.interval = 0.5
Sets the measure interval for SMU channel A to 0.5 s.
Also see
reset()
(on page 15-176)
setup.recall()
smuX.measure.count
smuX.reset()
(on page 15-231)
2606B-901-01 Rev. C November 2021
Affected by
Instrument reset
SMU reset
Recall setup
The interval value (in seconds); set to a value between 0 and 1
SMU channel (for example, smua.measure.interval applies to SMU Channel A)
(on page 15-191)
(on page 15-215)
Section 15: TSP command reference
Where saved
Default value
Saved setup
0 (0 s)
15-221