Section 15: TSP command reference
trigger.blender[N].clear()
This function clears the blender event detector and resets the overrun indicator of blender N.
Type
TSP-Link accessible
Function
Yes
Usage
trigger.blender[N].clear()
N
Details
This command sets the blender event detector to the undetected state and resets the overrun
indicator of the event detector.
Example
trigger.blender[2].clear()
Clears the event detector for blender 2.
Also see
None
trigger.blender[N].EVENT_ID
This constant contains the trigger blender event number.
Type
TSP-Link accessible
Constant
Yes
Usage
eventID = trigger.blender[N].EVENT_ID
eventID
N
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to
trigger events from this trigger blender.
Example
digio.trigger[1].stimulus = trigger.blender[2].EVENT_ID
Set the trigger stimulus of digital I/O trigger 1 to be controlled by the trigger blender 2 event.
Also see
None
15-362
Model 2606B System SourceMeter® Instrument Reference Manual
Affected by
Where saved
The blender number (up to six)
Affected by
Where saved
Trigger event number
The blender number (up to six)
Default value
Default value
2606B-901-01 Rev. C November 2021