Model 2606B System SourceMeter® Instrument Reference Manual
Also see
Adjustment
(on page 17-12)
reset()
(on page 15-176)
smuX.cal.lock()
smuX.cal.unlock()
smuX.measure.calibrateY()
smuX.reset()
(on page 15-231)
smuX.source.calibrateY()
smuX.cal.restore()
This function loads a stored set of calibration constants.
Type
TSP-Link accessible
Function
Yes
Usage
smuX.cal.restore()
smuX.cal.restore(calset)
X
calset
Details
This function overwrites the present set of calibration constants with constants read from
nonvolatile memory.
This function is disabled until a successful call to smuX.cal.unlock() is made.
If calset is not specified, smuX.CALSET_DEFAULT is used.
Example
smua.cal.restore()
Also see
Adjustment
(on page 17-12)
smuX.cal.lock()
smuX.cal.unlock()
2606B-901-01 Rev. C November 2021
(on page 15-199)
(on page 15-204)
(on page 15-214)
(on page 15-234)
Affected by
Source-measure unit (SMU) channel (for example, smua.cal.restore() applies
to SMU channel A)
The calibration set to be loaded; set calset to one of the following values:
▪
0 or smuX.CALSET_NOMINAL: A set of calibration constants that are
uncalibrated, but set to nominal values to allow rudimentary functioning of the
instrument
▪
1 or smuX.CALSET_FACTORY: The calibration constants when the instrument
left the factory
▪
2 or smuX.CALSET_DEFAULT: The normal calibration set
▪
3 or smuX.CALSET_PREVIOUS: The calibration set that was used before the
last default set was overwritten
(on page 15-199)
(on page 15-204)
Where saved
Restores factory calibration constants for SMU channel A.
Section 15: TSP command reference
Default value
15-201