Model 2606B System SourceMeter® Instrument Reference Manual
6. Change the subnet mask from the LAN CONFIG menu:
a.
Select CONFIG > SUBNETMASK, and then press the ENTER key. The SUBNETMASK menu item is to
the right of GATEWAY. Use the navigation wheel to scroll through the options.
b.
Modify the SUBNETMASK value to match the computer settings recorded earlier (or
255.255.255.000 if DHCP Enabled = YES).
c.
Press the ENTER key or the navigation wheel when you are finished changing all the characters.
d.
Press the EXIT (LOCAL) key twice to return to the LAN CONFIG menu.
7. From the LAN CONFIG menu, select APPLY_SETTINGS > YES, and then press the ENTER
key.
8. Open a web browser on the host computer.
9. Enter the IP address of the instrument in the web browser address box. For example, if the
instrument IP address is 192.168.1.101, enter 192.168.1.101 in the browser address box.
10. Press Enter on the computer keyboard to open the instrument web page.

LAN speeds

Another characteristic of the LAN is speed. The 2606B negotiates with the host computer and other
LXI-compliant devices on the LAN to transmit data at the highest speed possible. LAN speeds must
be configured to match the speed of the other instruments on the network.
To set the LAN speed:
1. From the virtual front panel, press the MENU key and select LAN > CONFIG > SPEED.
2. Turn the navigation wheel to select either 10 Mbps or 100 Mbps.
3. Press the ENTER key.
4. Press the EXIT (LOCAL) key once to return to the previous menu.
5. Select APPLY_SETTINGS > YES, and then press the ENTER key.

Duplex mode

The duplex mode is based on the LAN configuration. There are two settings:
Half-duplex: Allows communications in both directions, but only one direction is active at a time
(not simultaneously).
Full: Permits communications in both directions simultaneously.
To set the duplex mode:
1. From the virtual front panel, press MENU key and select LAN > CONFIG > DUPLEX.
2. Turn the navigation wheel to select either HALF or FULL.
3. Press the ENTER key.
4. Press the EXIT (LOCAL) key once to return to the LAN CONFIG menu.
5. Select APPLY_SETTINGS > YES, and then press the ENTER key.
2606B-901-01 Rev. C November 2021
Section 9: LAN concepts and settings
9-7