Tektronix KEITHLEY SourceMeter 2606B Reference Manual page 736

Table of Contents

Advertisement

Model 2606B System SourceMeter® Instrument Reference Manual
trigger.blender[N].orenable
This attribute selects whether the blender performs OR operations or AND operations.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
orenable = trigger.blender[N].orenable
trigger.blender[N].orenable = orenable
orenable
N
Details
This command selects whether the blender waits for any one event (OR) or waits for all selected
events (AND) before signaling an output event.
Example
trigger.blender[1].orenable = true
trigger.blender[1].stimulus[1] = digio.trigger[3].EVENT_ID
trigger.blender[1].stimulus[2] = digio.trigger[5].EVENT_ID
Generate a trigger blender 1 event when a digital I/O trigger happens on line 3 or 5.
Also see
trigger.blender[N].reset()
trigger.blender[N].overrun
This attribute indicates whether or not an event was ignored because of the event detector state.
Type
TSP-Link accessible
Attribute (R)
Yes
Usage
overrun = trigger.blender[N].overrun
overrun
N
2606B-901-01 Rev. C November 2021
Affected by
Where saved
Instrument reset
Not saved
Trigger blender N reset
Recall setup
The type of operation:
true: OR operation
false: AND operation
The blender number (up to six)
(on page 15-364)
Affected by
Where saved
Instrument reset
Not applicable
Trigger blender N clear
Trigger blender N reset
Trigger blender overrun state (true or false)
The blender number (up to six)
Section 15: TSP command reference
Default value
false (AND mode)
Default value
Not applicable