Model 2606B System SourceMeter® Instrument Reference Manual
To enable or disable DNS registration:
1. From the virtual front panel, press the MENU key and select LAN > CONFIG > DNS > DYNAMIC.
2. Turn the navigation wheel to select either ENABLE or DISABLE. DNS registration works with the
DHCP to register the host name specified in the lan.config.dns.hostname attribute with the
DNS server.
3. Press the ENTER key.
4. Press the EXIT (LOCAL) key twice to return to the LAN CONFIG menu.
5. Select APPLY_SETTINGS > YES, and then press the ENTER key.
To set the DNS server IP addresses:
1. From the virtual front panel, press the MENU key and select LAN > CONFIG > DNS.
2. Turn the navigation wheel to select either DNS-ADDRESS1 or DNS-ADDRESS2.
3. Press the ENTER key.
4. Turn the navigation wheel to select and enter a valid IP address for the DNS server.
5. Press the ENTER key.
6. Press the EXIT (LOCAL) key twice to return to the LAN CONFIG menu.
7. Select APPLY_SETTINGS > YES, and then press the ENTER key.
Establishing a point-to-point connection
To enable access to the instrument web page and other web applications from a computer, use a
one-to-one LAN connection and set up a static IP address between the host computer and
the instrument.
The following instructions describe how to configure the IP address of the instrument. The IP address
of the instrument is based on the present IP address of the host computer. Each device on the LAN
(corporate or private) requires a unique IP address.
Do not change the IP address at any time without talking to your system administrator.
Entering an incorrect IP address can prevent your workstation from connecting to your
corporate network.
If you have problems, see
2606B-901-01 Rev. C November 2021
LAN troubleshooting suggestions
Section 9: LAN concepts and settings
(on page 12-6).
9-5
Need help?
Do you have a question about the KEITHLEY SourceMeter 2606B and is the answer not in the manual?