Section 15: TSP command reference
Also see
Saved setups
setup.save()
(on page 15-192)
setup.save()
This function saves the present setup as a user-saved setup.
Type
TSP-Link accessible
Function
Yes
Usage
setup.save(id)
id
Details
When the id parameter is an integer (n), it is interpreted as the setup number to save to the
nonvolatile memory of the instrument.
When you save to a specified integer (1 to 5) in nonvolatile memory, the previous setup at that same
location is overwritten.
When the id parameter is a string, it is interpreted as the path and file name of the location to save
the present setup on a USB flash drive. The path may be absolute or relative to the current working
directory.
Example
setup.save(5)
Also see
Saved setups
setup.recall()
15-192
(on page 4-13)
Affected by
An integer or string specifying where to save the user setup:
▪
Save in nonvolatile memory (1 to 5)
▪
Save as user-saved setup on a USB flash drive ("/path/filename")
(on page 4-13)
(on page 15-191)
Model 2606B System SourceMeter® Instrument Reference Manual
Where saved
Saves the present setup to the internal memory of the
instrument at location 5.