Model 2606B System SourceMeter® Instrument Reference Manual
smuX.trigger.source.action
This attribute enables or disables sweeping the source (on or off).
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
action = smuX.trigger.source.action
smuX.trigger.source.action = action
action
X
Details
This attribute is used to enable or disable source level changes during a sweep. In addition to
enabling the action before initiating the sweep, make sure to configure it using
smuX.trigger.source.linearY(), smuX.trigger.source.listY(), or
smuX.trigger.source.logY().
Example
smua.trigger.source.listv({3, 1, 4, 5, 2})
smua.trigger.source.action = smua.ENABLE
Configure list sweep for SMU channel A (sweep through 3 V, 1 V, 4 V, 5 V, and 2 V).
Enable the source action.
Also see
smuX.trigger.source.linearY()
smuX.trigger.source.listY()
smuX.trigger.source.logY()
Triggering
(on page 7-1)
2606B-901-01 Rev. C November 2021
Affected by
SMU reset
Instrument reset
Recall setup
Sweep source action. Set to one of the following values:
▪
0 or smuX.DISABLE: Do not sweep the source
▪
1 or smuX.ENABLE: Sweep the source
Source-measure unit (SMU) channel (for example,
smua.trigger.source.action applies to SMU channel A)
(on page 15-265)
(on page 15-266)
(on page 15-267)
Section 15: TSP command reference
Where saved
Default value
Not saved
0 (smuX.DISABLE)
15-263