Section 15: TSP command reference
bufferVar.statuses
This attribute contains the status values of readings in the reading buffer.
Type
TSP-Link accessible
Attribute (R)
Yes
Usage
statusInformation = bufferVar.statuses[N]
statusInformation
bufferVar
N
Details
This read-only buffer recall attribute is like an array (a Lua table) of the status values for all the
readings in the buffer. The status values are floating-point numbers that encode the status value; see
the following table for values.
For dedicated reading buffers, all buffer attributes are saved to nonvolatile memory only when the
reading buffer is saved to nonvolatile memory.
Buffer status bits
Bit
Name
Overtemp
B1
AutoRangeMeas
B2
AutoRangeSrc
B3
4Wire
B4
Rel
B5
Compliance
B6
Filtered
B7
15-34
Model 2606B System SourceMeter® Instrument Reference Manual
Affected by
Where saved
Clearing the buffer
See Details
The status value when reading N of the specified buffer was acquired
The reading buffer; can be a dynamically allocated user-defined buffer or a
dedicated reading buffer
The reading number N; can be any value from 1 to the number of readings in the
buffer; use the bufferVar.n command to determine the number of readings in
the buffer
Hex
Description
0x02
Over temperature condition
0x04
Measure range was autoranged
0x08
Source range was autoranged
0x10
4-wire (remote) sense mode enabled
0x20
Relative offset applied to reading
0x40
Source function was limited because the complementary
function would be over the compliance limit
0x80
Reading was filtered
Default value
Not applicable
2606B-901-01 Rev. C November 2021