Tektronix KEITHLEY SourceMeter 2606B Reference Manual page 566

Table of Contents

Advertisement

Model 2606B System SourceMeter® Instrument Reference Manual
smuX.abort()
This function terminates all overlapped operations on the specified source-measure unit (SMU).
Type
TSP-Link accessible
Function
Yes
Usage
smuX.abort()
X
Details
The smuX.abort() function does not turn the output off or change any settings.
If this function is used to abort a sweep, when it is executed, the SMU exits its trigger model
immediately and returns to the idle state of the trigger model.
Example
smua.abort()
Also see
smuX.measure.overlappedY()
smuX.trigger.initiate()
smuX.buffer.getstats()
This function returns the statistics for a specified reading buffer.
Type
TSP-Link accessible
Function
Yes
Usage
statistics = smuX.buffer.getstats(bufferVar)
statistics
X
bufferVar
Details
This function returns a table with statistical data about the data that is placed in the buffer.
The SMU automatically updates reading buffer statistics as data is added to the reading buffer. When
the reading buffer is configured to wrap around and overwrite older data with new data, the buffer
statistics include the data that was overwritten.
The table returned from this function is a snapshot. Although the SMU continues to update the
statistics, the table returned is not updated. To get fresh statistics, call this function again.
2606B-901-01 Rev. C November 2021
Affected by
Source-measure unit (SMU) channel (for example, smua.abort() applies to SMU
channel A)
Terminates all overlapped operations on SMU channel A.
(on page 15-223)
(on page 15-256)
Affected by
The statistical data about the data in the reading buffer
Source-measure unit (SMU) channel (for example, smua.buffer.getstats()
specifies SMU channel A)
The reading buffer to process
Section 15: TSP command reference
Where saved
Default value
Where saved
Default value