Model 2606B System SourceMeter® Instrument Reference Manual
Set the instrument password
You can change the instrument password from the web interface. The instrument password is used
for access to the instrument from any remote interface. When password usage is enabled by the
localnode.passwordmode command, you must supply a password to change the configuration or
to control an instrument from a remote command interface.
When a password is set for the web interface and localnode.passwordmode is set to an option
that requires a password, you cannot make changes using the web interface options Virtual Front
Panel, Flash Upgrade, and TSB Embedded.
The default password is "" (no password).
You cannot retrieve a lost password from any command interface. You can reset the password from
the front panel. Select MENU, then RESET-PASSWORD.
For information on the TSP command, refer to
To change the password:
1. Select Set Password.
2. In the Current Password box, enter the presently used password.
3. In the New Password and Confirm New Password boxes, enter the new password.
4. Select Submit.
View buffer data using the web interface
You can view the data in the reading buffers on the Reading Buffers page of the Virtual Front Panel
web page.
To view the data, at the top of the virtual front panel, select Reading Buffers. On the Reading Buffers
page, select a reading buffer from the Reading Buffer list. The data is displayed in the table.
Initially, the reading number and reading measurement values are displayed. You can display
additional data using the filters to the right of the table. You can choose:
•
Measurement: Include or exclude measurement information.
•
Source: Include or exclude source information.
•
Other: Include or exclude status, status detail, and timestamp information. For detail on the
Status value, refer to
2606B-901-01 Rev. C November 2021
localnode.passwordmode
Buffer status
(on page 6-13).
Section 3: Instrument description
(on page 15-157).
3-19