Model 2606B System SourceMeter® Instrument Reference Manual
Details
If one or more trigger events have been detected since the last time lan.trigger[N].wait() or
lan.trigger[N].clear() was called, this function returns immediately.
After waiting for a LAN trigger event with this function, the event detector is automatically reset and
rearmed regardless of the number of events detected.
Example
triggered = lan.trigger[5].wait(3)
Wait for a trigger with LAN packet 5 with a timeout of 3 seconds.
Also see
lan.trigger[N].assert()
lan.trigger[N].clear()
lan.trigger[N].overrun
lan.trigger[N].stimulus
localnode.autolinefreq
This attribute enables or disables automatic power line frequency detection at start-up.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
flag = localnode.autolinefreq
localnode.autolinefreq = flag
flag
Details
When this attribute is set to true, the power line frequency is detected automatically the next time
the 2606B powers up. After the power line frequency is automatically detected at power-up, the
localnode.linefreq attribute is set automatically to 50 or 60.
If the localnode.linefreq attribute is explicitly set, localnode.autolinefreq is automatically
set to false.
When using this command from a remote node, replace localnode with the node reference, for
example node[5].autolinefreq.
Also see
localnode.linefreq
2606B-901-01 Rev. C November 2021
(on page 15-144)
(on page 15-144)
(on page 15-149)
(on page 15-151)
Affected by
Not applicable
The auto line frequency detection setting:
true: Enable automatic line frequency detection at start-up
false: Disable automatic line frequency detection at start-up
(on page 15-155)
Section 15: TSP command reference
Where saved
Default value
Nonvolatile memory
true (enabled)
15-153