Section 15: TSP command reference
Example
print(localnode.model)
Outputs the model number of the local node. For example:
2606B
Also see
localnode.serialno
localnode.password
This attribute stores the remote access password.
Type
TSP-Link accessible
Attribute (W)
Yes
Usage
localnode.password = "password"
password
Details
This write-only attribute stores the password that is set for any remote interface. When password
usage is enabled (localnode.passwordmode), you must supply a password to change the
configuration or to control an instrument from a remote command interface.
The instrument continues to use the old password for all interactions until the command to change it
executes. When changing the password, give the instrument time to execute the command before
attempting to use the new password.
You cannot retrieve a lost password from any command interface.
You can reset the password by resetting the LAN from the virtual front panel or by sending the
lan.reset() command.
When using this command from a remote node, localnode should be replaced with the node
reference, for example, node[5].password.
Example
localnode.password = "N3wpa55w0rd"
Changes the remote interface password to N3wpa55w0rd.
Also see
lan.reset()
(on page 15-135)
localnode.passwordmode
15-156
Model 2606B System SourceMeter® Instrument Reference Manual
(on page 15-161)
Affected by
LAN reset
LAN restore defaults
A string that contains the remote interface password, up to 255 characters
(on page 15-157)
Where saved
Default value
Nonvolatile memory
""
2606B-901-01 Rev. C November 2021
Need help?