Section 15: TSP command reference
reset()
This function resets commands to their default settings.
Type
TSP-Link accessible
Function
No
Usage
reset()
reset(system)
system
Details
The reset() command in its simplest form resets the entire TSP-enabled system, including the
controlling node and all subordinate nodes.
If you want to reset a specific instrument, use either the localnode.reset() or
node[X].reset() command. Use the localnode.reset() command for the local instrument.
Use the node[X].reset() command to reset an instrument on a subordinate node.
You can only reset the entire system using reset(true) if the node is the master. If the node is not
the master node, executing this command generates an error.
Example
reset(true)
If the node is the master node, the entire system is reset; if the node is not the master node, an error
is generated.
Also see
localnode.reset()
savebuffer()
This KISavebuffer factory script function saves a specified reading buffer as either a CSV file or an XML file.
Type
TSP-Link accessible
Function
No
Usage
savebuffer(buffer, "formatType", "fileName")
buffer
formatType
fileName
15-176
Affected by
What to reset:
▪
true: If the node is the master, the entire system is reset (default)
▪
false: Only the local group is reset
(on page 15-159)
Affected by
The reading buffer to save
A string indicating which file type to use: csv or xml
The file name of the saved buffer
Model 2606B System SourceMeter® Instrument Reference Manual
Where saved
Where saved
2606B-901-01 Rev. C November 2021
Default value
Default value
Need help?
Do you have a question about the KEITHLEY SourceMeter 2606B and is the answer not in the manual?