Model 2606B System SourceMeter® Instrument Reference Manual
Also see
localnode.showerrors
tsplink.reset()
localnode.prompts4882
This attribute enables and disables the generation of prompts for IEEE Std 488.2 common commands.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
prompting = localnode.prompts4882
localnode.prompts4882 = prompting
prompting
Details
When this attribute is enabled, the IEEE Std 488.2 common commands generate prompts if
prompting is enabled with the localnode.prompts attribute. If localnode.prompts4882 is
enabled, limit the number of *trg commands sent to a running script to 50 regardless of the setting
of the localnode.prompts attribute.
When this attribute is disabled, IEEE Std 488.2 common commands do not generate prompts. When
using the *trg command with a script that executes trigger.wait() repeatedly, disable
prompting to avoid problems associated with the command interface input queue filling.
Example
localnode.prompts4882 = 0
Disables IEEE Std 488.2 common command prompting.
Also see
localnode.prompts
localnode.reset()
This function resets the local node instrument.
Type
TSP-Link accessible
Function
Yes
Usage
localnode.reset()
2606B-901-01 Rev. C November 2021
(on page 15-161)
(on page 15-380)
Affected by
Power cycle
IEEE Std 488.2 prompting mode:
Disable prompting: 0
Enable prompting: 1
(on page 15-158)
Affected by
Section 15: TSP command reference
Where saved
Default value
Not saved
1 (enabled)
Where saved
Default value
15-159