Section 15: TSP command reference
smuX.cal.unlock()
This function enables the commands that change calibration settings.
Type
TSP-Link accessible
Function
Yes
Usage
smuX.cal.unlock(password)
X
password
Details
This function enables the calibration functions to change the calibration settings.
The password when the instrument is shipped from the factory is "KI0026XX".
Example
smua.cal.unlock("KI0026XX")
Also see
Adjustment
(on page 17-12)
smuX.cal.lock()
smuX.cal.password
smuX.cal.restore()
smuX.cal.state
smuX.contact.calibratehi()
This function adjusts the high/sense high contact check measurement.
Type
TSP-Link accessible
Function
Yes
Usage
smuX.contact.calibratehi(cp1Measured, cp1Reference, cp2Measured, cp2Reference)
X
cp1Measured
cp1Reference
cp2Measured
cp2Reference
15-204
Affected by
Source-measure unit (SMU) channel (for example, smua.cal.unlock() applies
to SMU channel A)
Calibration password
(on page 15-199)
(on page 15-199)
(on page 15-201)
(on page 15-203)
Affected by
Source-measure unit (SMU) channel (for example,
smua.contact.calibratehi() applies to SMU channel A)
The value measured by this SMU for point 1
The reference measurement for point 1 as measured externally
The value measured by this SMU for point 2
The reference measurement for point 2 as measured externally
Model 2606B System SourceMeter® Instrument Reference Manual
Where saved
Unlocks calibration for SMU channel A.
Where saved
2606B-901-01 Rev. C November 2021
Default value
Default value