31 Help and error messages
31.3.14 Water leak related error messages
31.3.15 Maintenance related errors
31.3.16 Software upgrade error messages
276
Planmeca Compact i
ActiveAqua related error messages
ERROR
EXPLANATION
CODE
E84.3
ActiveAqua water flow self-test fail.
Water flow not detected in ActiveAqua self-test. Check the
flow meter, circulation valve and pump.
Water leak related error messages
ERROR
EXPLANATION
CODE
E90
Water leak detected.
Main water valve is closed.
E90.1
Water leak detected. Water use allowed temporarily.
Maintenance related errors
ERROR
EXPLANATION
CODE
E91
Yearly maintenance overdue.
Yearly maintenance should have been performed <x> days
ago.
Software upgrade error messages
ERROR
EXPLANATION
CODE
E99
Software upgrade failure.
Retry upgrade.
E99.1
MCB software upgrade failure.
Retry upgrade.
E99.2
IMUX software upgrade failure.
Retry upgrade.
E99.3
Control panel software upgrade failure.
Retry upgrade.
E99.4
Operating light software upgrade failure.
Retry upgrade.
E99.5
Motorised headrest software upgrade failure.
Retry upgrade.