Keyboard Assembly Replacement Procedures

8-4-1
Keyboard Bottom Cover Replacement Procedure
NOTICE
8-4-1-1
Tools
Use the appropriate screwdriver, as indicated in the Keyboard Bottom Cover replacement procedure.
8-4-1-2
Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
page
4-7.
8-4-1-3
Keyboard Bottom Cover Removal Procedure
1.) Remove the Keyboard Assembly, as described in the
page
2.) Carefully place the Keyboard Assembly upside-down on a flat, clean, stable surface.
3.) Loosen and remove the 8 Phillips screws that secure the bottom cover in place - see
2
EVISION
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to
"Battery Removal Procedure" on page
8-39.
Figure 8-73 Keyboard Bottom Cover Securing Screws
Chapter 8 - Replacement Procedures
VS5 N
VS6 N
AND
8-116.
4-2-3 "Power Shut Down" on
"Keyboard Assembly Removal Procedure" on
S
M
ERVICE
ANUAL
Figure 8-73
8-59