P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
7.) Disconnect the up-down/swivel cable from the handle to be replaced and slide the handle sideways
to remove it from the supporting shaft -
8-3-7-4
Swivel and Up-Down Handle Installation Procedure
1.) Install the appropriate handle on the support shaft.
2.) Return spring/s and securing clip/s to secure the handle/s.
3.) Return the protective metal plate and secure it with the retaining screw (previously removed).
4) Install the Probe Shelf as described in the
5.) Install the Operator Panel as described in the
page
6) Lower the console to the maximum down position.
7.) Turn ON power to the system.
8.) Check for proper up/down and swivel movement of the handle, as described in
Functional Tests" on page
2
EVISION
Figure 8-69 Release Cable and Slide Handle Sideways
8-42.
4-35.
Chapter 8 - Replacement Procedures
VS5 N
Figure
8-69.
"Probe Shelf Installation Procedure" on page
"Keyboard Assembly Installation Procedure" on
VS6 N
S
M
AND
ERVICE
ANUAL
8-52.
"Mechanical
8-55