Probe Current Leakage Test
10-6-9-1
Definition
This test measures the current that would flow to Ground (from any of the probes) through a patient who
is being scanned, and who becomes grounded by touching some other grounded surface.
10-6-9-2
Generic Procedure
Measurements should be made with the Ground OPEN and CLOSED, with power line polarity Normal
and Reversed, and with the unit OFF and ON. For each combination, the probe must be active to find
the worst case condition.
H (BLACK)
N (WHITE)
POWER
OUTLET
G (GREEN)
NOTE:
Each probe will have a certain amount of current leakage, depending on its design. Small variations in
probe current leakages are normal from probe to probe. Other variations will result from differences in
line voltage and test lead placement.
2
EVISION
POLARITY REVERSING SWITCH
MOMENTARY
SWITCH
Figure 10-5 Set Up for Probe Current Leakage Test
Chapter 10 - Care and Maintenance
VS5 N
VS6 N
S
AND
ERVICE
PROBE
CONSOLE
LEAKAGE TEST
METER
M
ANUAL
10-29