P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
7-4-3-4
RFI Analog Circuits Test (cont'd)
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment
it fails. Alternatively, you may prefer testing to continue on failure and later you may review the
report.
5.) Trackball to the Start button and press Set.
The RFI Analog Circuits Test commences. While the test proceeds, the name of the test currently
in progress is displayed in the Execute field and the current Loop # and specific test are shown
below. The messages Init Done and Please Wait are displayed beside the progress bar in the
Status area:
As the testing sequence progresses, the progress bar will advance to reflect the test progress. The
LED on the RFI status indicator in the Data Flow map will be highlighted in the appropriate color to
indicate the current test status, as follows:
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the RFI Analog Circuits Test sequence is complete,
-
-
2
EVISION
Green: All recommended tests for these components have been completed and no problems
were identified.
Red:
Problems were identified in these components during the performance of the test.
Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
the Diagnostic Test window displays the Finished message
the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in
Analog Circuits Test" on page 7-40