P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
8-3-6
Probe Shelf Replacement Procedure
8-3-6-1
Tools
Use the appropriate screwdriver as indicated in the Probe Shelf Replacement procedure.
8-3-6-2
Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
page
4-7.
8-3-6-3
Probe Shelf Removal Procedure
1) Remove the Keyboard Assembly as described in the
page
2) Working from below, and while carefully supporting the Probe Shelf Holder, remove the 4 retaining
screws - two on either side - as seen in
Retaining Screw
Figure 8-65 Probe Shelf - Viewed from Below indicating Retaining Screws
3.) Remove the Probe Shelf.
8-3-6-4
Probe Shelf Installation Procedure
1.) Return the Probe Shelf to its location on the Keyboard Interface Assembly and secure it with the
4 retaining screws (previously removed).
2.) Install the Keyboard Assembly as described in the
page
3) Lower the console to the maximum down position.
4.) Turn ON power to the system.
5.) Check for proper up/down and side-to-side movement of the operator panel and monitor, as
described in
6.) Proceed to perform the following functionality tests:
-
-
8-52
Section 8-3 - Control Console Components Replacement Procedures
2
EVISION
8-39.
8-42.
"Mechanical Functional Tests" on page
"Grounding Continuity" on page 10-23
"Chassis Current Leakage Test" on page 10-24
"Keyboard Assembly Removal Procedure" on
Figure
8-65.
"Keyboard Assembly Installation Procedure" on
4-35.
VS5 N
VS6 N
S
AND
ERVICE
4-2-3 "Power Shut Down" on
Retaining Screw
M
ANUAL