GE Vivid S5 N Service Manual page 314

Hide thumbs Also See for Vivid S5 N:

Advertisement

P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
4-4-6
Probe/Connectors Check
TAKE THE FOLLOWING PRECAUTIONS WITH THE PROBE CABLES:
CAUTION
- KEEP AWAY FROM THE WHEELS
- DO NOT BEND
- DO NOT CROSS CABLES BETWEEN PROBES
Table 4-24
Step
1.
2.
3.
4.
5.
6.
7.
8.
4-44
2
EVISION
Probe and Connectors Checks
Task
Press Probe on the Operator Panel.
If not already selected. Use the trackball to select the desired
probe.
Trackball to the desired application
Press Select to launch the application.
To change application without changing the current probe,
press Appl. on the Operator Panel.
Verify no missing channels
Verify there are no EMI/RFI or artifacts specific to the probe. No EMI/RFI or artifacts.
Test the probe in each active connector slot.
See
'Connecting Probes" on page 3-71.
Do a leakage test on the probe.
See
'Electrical Safety Tests" on page 10-20.
Repeat this procedure for all available probes.
Section 4-4 - Image Testing
VS5 N
VS6 N
AND
Expected Result(s)
A list of the connected probes will pop up on the screen.
An application menu for the desired probe is listed on
the screen.
The selected application starts.
All channels are functioning.
The screen will display pictorial data each time
The connected probe passes the test.
S
M
ERVICE
ANUAL