Download Print this page

GE Vivid S5 N Service Manual page 606

Hide thumbs Also See for Vivid S5 N:

Advertisement

P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
Replacement
On/Off Switch
Replacement
Screen
Sensor
Figure 8-94 Extended PCB Keyboard (Top left: spare On/off switch and screen sensor)
8-4-8-1
Tools
Use the appropriate flat and Phillips-type screwdrivers, Hex wrench keys, and a No. 10 Torx
screwdriver.
8-4-8-2
Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
page
4-7.
8-4-8-3
Extended Keyboard Removal Procedure (New-type Operator Panel)
1.) Remove the trackball from the Keyboard Assembly.
2.) Remove the Keyboard Assembly, as described in the
page
3.) Remove the Keyboard Bottom Cover, as described in the
Procedure" on page
4.) Uninstall trackball housing as described in the
Procedure" on page
5.) Disconnect the seven cables - refer to the illustration in
8-80
Section 8-3 - Control Console Components Replacement Procedures
2
EVISION
8-39.
8-59.
8-71.
VS5 N
4-2-3 "Power Shut Down" on
"Keyboard Assembly Removal Procedure" on
"Keyboard Bottom Cover Removal
"Mechanical Trackball and Housing Removal
Figure 8-97 on page
VS6 N
S
M
AND
ERVICE
ANUAL
8-83.

Hide quick links:

Advertisement

loading

This manual is also suitable for:

Vivid s6 n