P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
8-3-1
Monitor Replacement Procedure
7.) Using a screwdriver, release the DVI cable and the power cable connectors from the monitor
(loosening the two screws on each side of the cables to disconnect them) - see
Power Cable
8.) Loosen and remove the four Phillips screws that secure the monitor to the support column, as
indicated in
Do not let the screws fall while performing this step
9.) Using two hands, carefully lift the monitor away from the securing bracket on the support column to
remove it from the system.
10.)Carefully lie the monitor face-down on a clean, soft surface.
8-28
Section 8-3 - Control Console Components Replacement Procedures
2
EVISION
Figure 8-30 Disconnecting the Power and DVI Cables
Figure 8-31
below.
Figure 8-31 Removing the Securing Screws