Cover Replacement Procedures - GE Vivid S5 N Service Manual

Hide thumbs Also See for Vivid S5 N:

Advertisement

P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
8-2-3
Left Side Cover Replacement Procedure
NOTE:
The left side cover is fastened in position on the side of the system by 4 ball and snap-lock securing clips.
This cover contains an air filter, held in position by securing clips.
For instructions on replacing the air filter only, see
8-2-3-1
Tools
None.
8-2-3-2
Preparations
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
page
4-7.
8-2-3-3
Left Side Cover Removal Procedure
1) Gripping the recess located near the bottom of the cover (as shown in
of the cover out towards you to release it from the lower securing clip.
2) Pull the top of the cover out towards you, releasing it from the upper securing clip.
3) Remove the left side cover (together with the attached air filter) from the system.
8-6
2
EVISION
Figure 8-4 Removing the Left Side Cover
Figure 8-5 Left Side Cover Removed
Section 8-2 - Cover Replacement Procedures
VS5 N
"Air Filter Replacement Procedure" on page
4-2-3 "Power Shut Down" on
Grip cover here
VS6 N
S
M
AND
ERVICE
ANUAL
8-7.
Figure
8-4), pull the bottom