Download Print this page

GE Vivid S5 N Service Manual page 331

Hide thumbs Also See for Vivid S5 N:

Advertisement

P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
5-3-3
RFI_AM Board
Table 5-3 RFI_AM Board - Input DC Voltages
+3.3 V
HVP_IN
HVN_IN
+THV
-THV
+5 V_IN
-5 V_IN
+12 V_IN
1.8 V
5-3-3-7
Probe Management
5-3-3-7-1
Mechanism
The RFI Board performs the following probe management functions:
Controls the selection of a probe connector
Senses the Probe type
Senses the Scan Plane Angle on TE probes
Senses the Probe temperature
If the temperature exceeds the maximum temperature, the RBI stops scanning and disables
Tx voltage.
The temperature references are set to 41.3
If the temperature reading falls below minimum temperature, the RBI stops scanning and disables
Tx voltage.
The minimum temperature references are set to 15
indicates a probe temperature-sensing error.
2
EVISION
(cont'd)
Input
Description
DC voltage from PS
Fixed DC high voltage +80 V
Fixed DC high voltage -80 V
Transmit high voltage
varies from 0 to +50 V
Transmit high voltage
varies from 0 to -50 V
DC voltage from PS
DC voltage from PS
DC voltage from PS
DC voltage from PS
Chapter 5 - Components and Function (Theory)
VS5 N
Connector - PIN#
Connected From
P14 1-4
From PS
P14-9
From PS
P14-11
From PS
P14-12
From PS
P14-14
From PS
P14-16
From PS
P14-20
From PS
P14-18
From PS
P1421, 22
From PS
o
o
C and 43.0
C
o
C. A temperature reading as low as 15
VS6 N
S
M
AND
ERVICE
ANUAL
o
C
5-13

Hide quick links: