Current Leakage Limits
The following limits are summarized for NFPA 99 (for USA) and IEC 60601-1 Medical Equipment Safety
Standards. These limits are GE standards and in some cases are lower than the above standards listed.
Table 10-13 Chassis Current Leakage Limits - Accessible Metal Surfaces
Country
Table 10-14 Type BF Applied Part Current Leakage Limits
Country
ALL
a. Single fault condition.
Table 10-15 Type CF Applied Part Current Leakage Limits - Surgical Probes and ECG Connections
Country
ALL
a. Single fault condition.
2
EVISION
Normal Condition
N/A
USA
0.1 mA
Other
- Non-Conductive (Floating) Surface and Cavity Probes
Normal
Condition
0.1 mA
Normal
Condition
0.01 mA
Chapter 10 - Care and Maintenance
Open Ground
0.3 mA
0.5 mA
Open
Reverse
Ground
Polarity
0.5 mA
0.1 mA
Open
Reverse
Ground
Polarity
0.05 mA
0.01 mA
VS5 N
VS6 N
S
AND
Reverse Polarity
Open Neutral
0.3 mA
0.5 mA
Open
Neutral
0.5 mA
Open
Neutral
0.05 mA
M
ERVICE
ANUAL
N/A
0.5 mA
Mains
a
Applied
5.0 mA
Mains
a
Applied
0.05 mA
10-21