eVue Dataflow Configuration
NOTE:
MPEGVue Player is intended for secondary review and patient consultation.
3-7-8-1
Introduction
It is possible to configure the Vivid S5 N or Vivid S6 N ultrasound scanner to allow the user to send
images or loops to a remote PC, in almost real time. When pressing Store, the loop will save not only
to the local archive (in raw-data format), but will also be transmitted to a designated PC ("e-Vue PC")
on the network (in compact MPEG format), and will appear at the review PC within a very short time.
This feature proves very useful in the OR scenario, where consultation with another doctor in the
hospital in real-time is required.
3-7-8-2
Preparations
Before configuring the Vivid S5 N or Vivid S6 N for eVue dataflow, it is necessary to decide upon a
specific PC in the hospital as being the e-Vue PC station. The service technician then configures the
Vivid S5 N or Vivid S6 N to send images to that specific e-Vue PC, as described below.
3-7-8-3
Configuring the Vivid S5 N or Vivid S6 N for eVue Dataflow
1.) Verify which PC in the hospital/institute has been designated as the e-Vue station.
2.) Make sure the Vivid S5 N or Vivid S6 N is connected to the Network via cable or wireless.
3.) On the Vivid S5 N or Vivid S6 N scanner, press the Config button.
The System Configuration window opens.
4.) Click Connectivity.
The Connectivity dialog box opens.
5.) Select the Dataflow Tab, as shown in
2
EVISION
Chapter 3 - System Setup
Figure 3-163 on page
3-142.
VS5 N
VS6 N
S
AND
ERVICE
M
ANUAL
3-141