8-6-2
Battery Replacement Procedure
NOTE:
This procedure is applicable to Vivid S6 N systems only.
8-6-2-1
Tools
Use the appropriate Phillips-type screwdrivers as indicated in the Battery replacement procedure.
8-6-2-2
Time Required
1 minute
8-6-2-3
Preparations
Shut down the Vivid S6 N ultrasound unit, as described in
8-6-2-4
Battery Removal Procedure
1) Remove the Vivid S6 N left and right side covers, as described in the
Procedure" on page 8-6
2) Remove the Vivid S6 N front cover, as described in the
17.
3) Open and remove the cabinet cage assembly, as described in
Cage Assembly" on page
4) Using a small Phillips screwdriver, loosen (but do not remove) the screw on the side of the battery
compartment (near the bottom of the cage - see
8-116
2
EVISION
and
Figure 8-144 Loosening the Screw on the Battery Compartment
Section 8-6 - Lower Section Components Replacement Procedures
"Right Side Cover Removal Procedure" on page
8-109.
Figure
VS5 N
AND
4-2-3 "Power Shut Down" on page
"Left Side Cover Removal
"Front Cover Removal Procedure" on page 8-
"Opening and Removing the Cabinet
8-144).
VS6 N
S
M
ERVICE
ANUAL
4-7.